OPERATING INSTRUCTIONS
Turning the T40 Rattler ON/OFF
On/Off
A
warm up routine which includes tests of the display,
audible, visual, and vibrating alarms. Following this
brief warm up, the T40 Rattler enters the gas readings
mode signified by the four digit display and gas
identifier icon. To turn off the T40 Rattler, simply press
and hold 'On/Off' for five seconds (four beeps).
Zeroing / Calibrating the T40 Rattler
B
C
D
The three status conditions are: GOOD – the sensor is healthy and should operate
for an extended period of time; PASS – the sensor passed calibration but had a
marginal response; or FAIL – the sensor did not pass calibration, meaning the
sensor has reached the end of its useful life, or the calibration gas source was not
accurate or has run out. Note: The T40 Rattler is factory programmed to accept 100
ppm (CO) and 25 ppm (H
S) span gas inputs. Gases in other concentrations will
2
result in inaccurate or failed calibrations.
1-800-DETECTS
412-788-4353
TM
The T40 Rattler is a single gas monitor designed to alert the user to rising concentrations
of either Carbon Monoxide (CO) or Hydrogen Sulfide (H
continuously display the ambient concentration of the designated gas and activate its
audible, visual and vibrating alarms whenever the preset limits are exceeded.
To turn the unit on, press and hold the 'On/Off'
button until the word "On" appears on the
display.
The T40 Rattler will power up and begin a
A function or "bump" test, using a known
concentration of calibration gas, should be
performed periodically based on instrument use,
exposure to gas, and environmental conditions.
The frequency is best determined by company
policy or local regulatory agencies. If an instru-
ment fails a function or "bump" test, or, if it is
dropped, submerged, or appears damaged, a full
calibration is recommended.
The safest approach is to perform a function or
"bump" test prior to each day's use. Industrial
Scientific is not responsible for establishing
customer safety practices and policies.
To zero the T40 Rattler, press
On/Off
Mode
'On/Off' and 'Mode' simultaneously
(C). The T40 Rattler will automati-
cally zero and beep after approxi-
mately 10 seconds, signaling that the zeroing
process is complete. If you do not wish to
perform a span calibration, flip the calibration
adapter back to its original position and the T40
Rattler will return to the gas reading mode.
Note: Never zero the T40 Rattler in an environ-
ment that is known to contain a background of
CO or H
S. For these environments, use zero air
2
gas.
After the zeroing process is complete, the T40
Rattler will beep, signaling it is ready to be span
calibrated. To perform a span calibration,
connect the supplied tubing between the
regulator and calibration adapter as shown (D),
turn the regulator's flow control to "open," and
press 'On/Off' and 'Mode' simultaneously.
The T40 Rattler will begin to show its response
to the applied gas as a full span on the display.
After approximately 1.5 minutes, the T40 Rattler
will finish the calibration routine and display the
status of the sensor. Remove tubing and flip
calibration adapter back to its original position.
www.indsci.com
Alarm Conditions / Operational Modes
The T40 Rattler is equipped with audible, visual, and
vibrating alarms to alert you when the ambient gas
concentration exceeds one of the two alarm levels (low
and high). For CO, the T40 Rattler is preset to alarm at
35 and 70 ppm. For H
10 and 20 ppm. To change these low and high settings,
see 'Setting the Configuration' below.
At any time during operation, you can access the peak
reading and alarm set point screens by pressing
'Mode.' The peak reading screen displays the highest
concentration of gas detected since the peaks were last reset. To reset
the peak reading, press 'On/Off.' The peak screen will reset to current
reading at the time of reset.
Pressing 'Mode' once more will show you the low alarm set point.
Pressing 'Mode' again will show you the high alarm set point. Pressing
'Mode' again will return you to the gas readings mode.
Changing the Batteries
The T40 Rattler operates for up to 1,500 hours on a
single "AA" alkaline battery (additional replacement
battery included). The icon on the display continuously
shows battery life. The T40 Rattler will alert you of a
low battery condition by emitting a beep once every 30
seconds. To change the battery, simply insert the
supplied hex tool into the battery cap, rotate 1/8 turn
counterclockwise, and remove battery (E). Install
new battery, being sure to observe proper polarity.
Replace the battery cap and turn clockwise with
Warning: Replace battery only in non-hazardous location.
Substitution of components may impair intrinsic safety.
Warranty
The T40 Rattler carries a two-year warranty from the date of
E
shipment. See included warranty certificate for coverage
details.
Setting the Configuration
Press and release the 'On/Off' and 'Mode' keys simultaneously during
the warm up routine when the countdown reaches between 15 and 10
seconds. Four options will be available to configure: High alarm value,
Low alarm value, Calibration gas concentration and Display. Press the
'On/Off' key to change the displayed value. Use the 'Mode' key to adjust
the value. Once the desired value is met, press the 'On/Off' key to accept
the value. If no buttons are pressed, the T40 will scroll to the next item.
When the display configuration is turned off, the T40 will display Safe,
Low, and High conditions instead of actual PPM levels.
After going through all configurations, the T40 will return to the warm up
routine. All changes are saved to memory when the unit is turned off.
Approvals
UL and cUL – Class I, Groups A, B, C, D T4 Ex ia
CSA – Ex ia IIC T4
Australia – ANZEx Ex ia IIC T4
CE-Mark (ATEX) –
IECEx- TSA 08.0009X Ex ia IIC T4 Ga
1-800-788-8383
412-788-8353
S). The T40 Rattler will
2
S, the preset alarm levels are
2
hex tool. Use only Rayovac LR6 or Duracell
MN1500
II 1G, DEMKO 02 ATEX 021349X Ex ia IIC T4 Ga
1 Life Way Pittsburgh,
Pennsylvania 15205-7500

Publicité

loading

Sommaire des Matières pour Industrial Scientific Rattler T40

  • Page 1 The T40 Rattler is a single gas monitor designed to alert the user to rising concentrations of either Carbon Monoxide (CO) or Hydrogen Sulfide (H S). The T40 Rattler will continuously display the ambient concentration of the designated gas and activate its OPERATING INSTRUCTIONS audible, visual and vibrating alarms whenever the preset limits are exceeded.
  • Page 2 Valor alto de la alarma, valor bajo de El T40 Rattler está equipado con alarmas auditiva, visual y Industrial Scientific recomienda que se efectúe una prueba la alarma, concentración del gas de calibración y Pantalla. Oprima vibratoria para alertarlo cuando la concentración de gas periódica de funcionamiento del T40 Rattler usando una...
  • Page 3 The company Industrial Scientific Corporation, Pittsburgh, Pennsylvania USA, declares that the following new material intended for use in Explosive Atmospheres: (La société Industrial Scientific Corporation, Pittsburgh, Pennsylvania USA, atteste que le matériel neuf destiné à être utilisé en Atmosphères Explosives désigné ci-après:) Gas detector T40 Rattler (Détecteur de gaz T40 Rattler)