Télécharger Imprimer la page

Agilent Technologies TV 301-P Navigator Notice De Mode D'emploi page 211

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 33
All manuals and user guides at all-guides.com
28
1
Sccm
29
1
-
TV 301-P Navigator Controller User Manual / 87-901-001-01 (B)
B1 - Pump Over-temperature
("Pump Bearing Temperature" or "Pump Body
Temperature" are over the limit)
B0 – Check Connection To Pump
(the connection between controller and pump missing)
Flow Meter Reading
(Specific option, not available on standard models)
B0 – Start/Stop Status
0 = Stop
1 = Start
B1 – Low Speed Status
(Read the "Low Speed" function status (Out Buff byte-0 B6))
0 = off (the "target speed"(In Buff byte-8)
is" high speed" (Out. Buff. byte-1))
1 = on (the "target speed" (In Buff byte-8)
is "low speed" (Out. Buff. byte-3))
B2 – Soft Start Status
0 = on
1 = off
(this function is active at first pump run-up and can't be
deactivated by the user, after first run-up activate this
function only if the pump has been stopped for at least 3
months; this procedure will take 1h)
B3 – Vent Valve Status
0 = close
1 = open
B4 – Purge Valve Status
Not used
B5 – Set Point Status
0 = open
1 = closed
B6 – Flow meter Alarm Status
Specific option, not available on standard models
B7 – Not Used
Technical Information
Profibus message mapping
17
223
000
001
100
122
145
221
212
211/224

Publicité

loading