Télécharger Imprimer la page

Agilent Technologies TV 301-P Navigator Notice De Mode D'emploi page 206

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 33
17
Technical Information
Profibus message mapping
Profibus message mapping
Table 10
Parameter Buffer (17 bytes)
Offset
Size
Unit res
byte
byte
0
1
°C
1
1
°C
2
1
%
3
1
°C
4
1
Sccm
5
2
7
2
Hz/mA/s
9
2
s
11
2
0,2*s
13
2
0,2*s
15
1
Bit Field
206/224
All manuals and user guides at all-guides.com
Range
Description
15-35
Water Cooling Max Temperature
(not used – insert 20)
1-5
Water Cooling Hysteresis
(not used – insert 2)
1-99
Setpoint Hysteresis
Set the hysteresis on set point output (see set point out
paragraph)
-
Heather Jacket Temperature Setting
Specific option, not available on standard models
-
Flowmeter Threshold
Specific option, not available on standard models
-
Reserved
-
Setpoint Value
Set the threshold for the set point out
0-999999
Setpoint Mask Time
Set the time during the controller does not drive the set
point out
0-65535
Vent Valve Opening Delay
0 = inf.
Set the delay time between the stop command and vent
opening.
Valid only in "purge/vent auto mode" (Parameter Buffer
byte-15 B1=0); default = 15; 0 = infinite (the valve will
never open)
0-65535
Vent Valve Open Time
0 = inf.
Set how long the vent will remain open, then both purge
and vent will be closed. Valid only in "purge/vent auto
mode" (Parameter Buffer byte-15 B1=0); 0 = infinite (the
valve remain open till a start command)
B0 - Vent & Purge Controlled Vent Procedure/Man
0 = Controlled Vent Procedure
1 = Manual (see Par. Buff. byte-15-B1)
TV 301-P Navigator Controller User Manual / 87-901-001-01 (B)
Win
serial
158
159
105
113
115
-
102
103
126
147
125

Publicité

loading