All manuals and user guides at all-guides.com
Installation
The ProfiBus connector pin-out is reported in the following figure.
Please in order to obtain best perform-ance use ProfiBus certified
cables.
To maintain the IP-54 protection level, use the appropriate connector provided
CAUTION!
with the pump.
Figure 18
Configuration
Before power-on the controller set the Profibus address by the 2
rotary switches.
Up to 126 different addresses (from 0 to 125) can be selected. The
address value sets by the 2 switches is expressed in hexadecimal
notation, so value from 00 to 7D are permitted; this setting is read by
the interface during the power-on phase, so any change of the switch
position after the power-on is ignored.
Before power-on the controller place the special mating connector on the serial
NOTE
connector J6 in order to ensure the communication between the internal link
and the Agilent turbo controller.
TV 301-P Navigator Controller User Manual / 87-901-001-01 (B)
ProfiBus connector
Technical Information
Profibus
201/224
17