6.2 METROLOGICAL CHARACTERISTICS

- Reference conditions

Quantity of influence
Ambient temperature
Relative humidity
Ext. magnetic field
Electric field
Battery voltage
Use in the reference range
- Optical sensor: matte target (one that produces no measurement in rotation without reflecting adhesive).
- External connector: standardized 0 - 5V TTL signal.
- RPM function
Range * rpm
6.0000
9.9999
Resolution
0.0006tr/min
Accuracy
Measurement time 11s > t > 7s
Stability
*
from 6 to 60rpm: usable only with the external input. Use up to 10,000rpm with the mechanical adapter.
- HZ function
At frequencies above 10kHz (upper limit of the range of use), the device may indicate erroneous values.
Range * Hz
0.1000
9.9999
Resolution
0.0004Hz
Accuracy
Measurement time 11s > t > 0,5s
Stability
*
from 0,1 to 1Hz : usable only with the external input.
Reference conditions
23°C
45% to 75% HR
< 40A/m to 50 or 60Hz
< 1V/m to 50 or 60Hz
9V
10.000
-
-
99.999
0.0006tr/min
1.10
-4
of reading ± 6 points
7s > t > 1s
10.000
-
-
99.999
0.004Hz
4.10
of reading ± 4 points
-5
± 4 points
53
53
100.00
1000.0
-
-
999.999
9999.9
0.0006tr/min
0.0006tr/min
1s > t > 1s
< 0,5s
± 6 points
100.00
1000.0
-
-
999.99
9999.9
0.04Hz
0.4Hz
< 0,5s
Tolerances
± 3k
± 0,5V
10000
-
99999
0.0006tr/min