4.4 Drying and storage

• Place all parts on a dry, clean and absorbent surface and let them dry completely.
• Wrap all the individual parts in a clean, lint-free cloth (e.g., a tea-towel) and keep in
a dry, dust-free environment.
NOTICE
Leaking batteries can cause damage to the device. Therefore, always remove
batteries or rechargeable batteries if you do not expect to use the device for a pro-
longed period [see: Inserting and replacing batteries, page 40].
5

TROUBLESHOOTING

Fault
The VELOX cannot be
switched on.
The VELOX is not
nebulising or has
unexpectedly stopped
nebulising.
Nebulising takes longer
than usual.
When cleaning with the
VELOXcare, not all of the
cleaning fluid was
circulated.
In the event of faults that are not listed in this chapter, or if the suggested procedure
does not correct the fault, contact the PARI GmbH Service Center.
– 55 –
Possible cause
The batteries are flat.
The power adapter is not
plugged into a socket
correctly, or the USB
connector is not properly
seated in the USB port on the
controller.
No medication has been
added.
The inhalation session has
been interrupted.
The maximum operating time
of 15 minutes per application
has elapsed.
Saline solution or medication
has got into the plug area.
The aerosol head is blocked.
The device switched off
automatically after 15 min.,
even though not all of the
liquid had passed through.
Procedure
Insert new batteries or connect
the power adapter.
Ensure that the power adapter
is plugged into the power
socket and the USB connector
is seated correctly in the
controller.
Add a suitable medication.
Press the On/Off button to
resume the inhalation session.
Rinse the nebuliser connector
with tap water and shake it
thoroughly to remove excess
water. Wipe a dry cloth over
the nebuliser connector on the
controller.
Rinse the aerosol head and
the VELOXcare.
Switch the controller on again
with the On/Off button. The
remaining fluid will then be
flushed through.