11 Troubleshooting

The scale should be switched off for a short time following an interruption in the pro-
gram sequence and disconnected from mains supply. It is then necessary to repeat
the weighing process from the beginning.
Help:
Interruption
Weight display is not illumi-
nated.
Weight display changes
continuously
The weighing result is obvi-
ously incorrect
Should other error messages occur, switch the scale off and then on again. If the
error message remains, inform the manufacturer.
12
Possible cause
• The scale is not switched on.
• Batteries are incorrectly inserted or discharged
• No batteries.
• Draught/air movement
• Table/floor vibrations
• The weighing plate is in contact with foreign mat-
ters or is installed incorrectly.
• Electromagnetic fields/static charging (choose a
different location for the scale, switch off an inter-
fering device if possible)
• The scale display is not set to zero
• Incorrect adjustment.
• Great fluctuations in temperature .
• Warm-up time was ignored.
• Electromagnetic fields/static charging (choose a
different location for the scale, switch off an inter-
fering device if possible)
MBE-BA-e-1612