9
Troubleshooting
The balance should be switched off for a short time following an interruption in the pro-
gramme sequence and disconnected from the mains supply. It is then necessary to repeat
the weighing process from the beginning.
Help:
Interruption
Weight display is not illuminated.
The weight display changes continu-
ally
The weighing result is obviously in-
correct
Switch the balance off if other error messages should appear and then switch on again.
Contact the manufacturer if the error message does not disappear.
572/573/KB/DS/FKB/KBJ-BA-e-1774
Possible cause
• The balance is not switched on.
• The mains supply connection has been inter-
rupted (mains cable not plugged in/faulty).
• Power supply interrupted. .
• Draught/air movement
• Table/floor vibrations
• The weighing plate is in contact with foreign
matter.
• Electromagnetic fields / static charging
(choose different location/switch off interfering
device if possible)
• The balance display is not set to zero
• Adjustment is no longer correct.
• Great fluctuations in temperature.
• Electromagnetic fields / static charging
(choose different location/switch off interfering
device if possible)
41