Wonderxcart 092; Area (M²) - Mercatus WonderXCart 092 Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

User's Manual
Manal de Instruções
Manual de Instrucciones
Manual mode
9.2.
WonderXCart maximum cycle time runs for an overall of 60 min. Use the MANUAL MODE to program personalized running times. We
recommend a safety tolerance of + 5 % to ensure maximum effectiveness, see table 3.
Table 3: Recommended cycle time, per dosage intensity per area for WonderXCart.

WonderXCart 092

Dosage (mJ/cm²)
Area
(m²)
1 min
5 min
15 min
< 10
4,4
40,0
128,8
10 ~ 15
3,0
26,6
85,8
15 ~ 20
2,2
20,0
64,4
20 ~ 25
1,8
16,0
51,5
25 ~ 30
1,5
13,3
42,9
30 ~ 35
1,3
11,4
36,8
35 ~ 40
1,1
10,0
32,2
40 ~ 45
1,0
8,9
28,6
45 ~ 50
0,9
8,0
25,8
Tools mode
9.3.
The TOOLS screen gives information regarding the unit alarms, lamps and allows to change the language.
Alarms
9.3.1.
All the alarms displayed in the ALARM screen automatically shut of the WonderXCart has a safety measure
For any maintenance check we recommend the use of safety gloves. Skin fat may interfere with lamps performance and
equipment reflection ability.
Table 4: WonderXCart controller alarms.
Problem
Alarm
No problem
LAMPs don't
turn on
Cycle doesn´t
starts
Cycle stopped
LAMPs
9.3.2.
The LAMPs STATUS screen displays the approximated life cycle of each lamp. When life cycle is below 500 hours, it will be blinking.
Press SET on each lamp to see the LAMP x counter screen. Check number of hours used, number of working hours remaining and a
reset counter button. Only reset counter after a lamp replacement (more in 12.2.1. Lamp counter reset)
Language
9.3.3.
Choose between the six available languages: English, French, German, Italian, Portuguese and Spanish. Press SET to select.
Notice d'Utilisation
Bedienungsanleitung
30 min
45 min
60 min
262,0
395,2
528,4
174,6
263,4
352,2
131,0
197,6
264,2
104,8
158,1
211,3
87,3
131,7
176,1
74,8
112,9
151,0
65,5
98,8
132,1
58,2
87,8
117,4
52,4
79,0
105,7
Possible Cause
Corrective Action
No Alarms
No corrective action
LAMP x life cycle
Replace LAMP X ; Unit will display alarms when lamp life cycle has left 500 h,
ended
250 h, 100 h, 50 h, 25 h and 10 h.
LAMP x faulty,
damaged or
Carefully check if lamp needs replacement or maintenance
disconnected
Motion sensor not
Unplug power from socket and replace motion sensor
operating
Movement detected
Enter the treatment room and press SET. The countdown for evacuation will
inside treatment room
start; 90 seconds to leave the room
I0001-2021-MNS-MKT
WonderXCart 113
Area
Dosage (mJ/cm²)
(m²)
1 min
5 min
15 min
< 10
6,4
57,2
184,4
10 ~ 15
4,2
38,2
123,0
15 ~ 20
3,2
28,6
92,2
20 ~ 25
2,5
22,9
73,8
25 ~ 30
2,1
19,1
61,5
30 ~ 35
1,8
16,4
52,7
35 ~ 40
1,6
14,3
46,1
40 ~ 45
1,4
12,1
41,0
45 ~ 50
1,3
11,4
36,9
30 min
45 min
60 min
375,2
566,0
756,8
250,2
377,4
504,6
187,6
283,0
378,4
150,1
226,4
302,7
125,1
188,7
252,3
107,2
161,7
216,2
93,8
141,5
189,2
83,4
125,8
168,2
75,0
113,2
151,4
(
.
table 4)
9 / 47

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Wonderxcart 113

Table des Matières