Electromagnetic immunity
The Naída CI is intended for use in the electromagnetic
environment specified below. The customer or the user
of the Naída CI should assure that it is used in such an
environment.
Immunity
Electrostatic
discharge
(ESD) IEC
61000-4-2
Power
frequency
(50/60 Hz)
Magnetic
Field
IEC 61000-
4-8
IEC
Compliance
60601
level
test level
± 6 kV
± 6 kV
contact
contact
± 8 kV air
± 8 kV air
3 A/m
3 A/m
Electromagnetic
environment -
a
guidance
Floors should be
wood, concrete
or ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30 %. As with the
operation of other
electronic devices,
precaution should
be taken to not
generate ESD.
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in
a typical
commercial or
hospital
environment.
21