Nice MhouseKit CL2SL Instructions Et Avertissements Pour L'installation Et L'utilisation page 45

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

9.6.2 - Flashing light
During the manoeuvre the flashing indicator flashes once every second.
When something is wrong the flashes are more frequent (every half sec-
Quick flashes
Status
ECSbus error
1 flash
1 second pause
1 flash
Photocell activated
2 flashes
1 second pause
2 flashes
"Gearmotor force" limiting device activated
3 flashes
1 second pause
3 flashes
Activation of the STOP input
4 flashes
1 second pause
4 flashes
9.6.3 - Control unit
The control unit has a series of LEDs, each of which can give particular
signals both during regular operation and when there is a problem, see
Table 13.
LED ECSBus [A]
Off
On
One flash per second
2 long flashes
1 flash every 5 seconds
Series of flashes separated
by a pause
Quick flash
LED STOP [B]
Off *
On
LED OPEN [C]
Off
On
TABLE 13 (fig. 27)
Status
Fault
Serious fault
All OK
Input status variation
Automation in "standby" mode
This is the same signal as on the flasher, see Table 12
ECSBus short circuit
Status
Activation of the STOP input
All OK
Status
All OK
OPEN input activated
ond); the light flashes twice with a second's pause between flashes, see
Table 12.
TABLE 12
Action
At the beginning of the manoeuvre the devices present do not cor-
respond with those recognised; check and eventually carry out the
learning procedure (9.3.3 "Recognition of other devices"). One or
more devices may be faulty; check and, if necessary, replace them.
At the start of the manoeuvre, one or more photocells do not ena-
ble movement; check to see if there are any obstructions. During
the movement if there is an obstruction no action is required.
During the movement, the gate experienced excessive friction;
identify the cause.
At the start of the manoeuvre or during the movement, the STOP
input was activated; identify the cause.
27
Action
Make sure there is power supply; check to see if the fuses are
blown; if necessary, identify the reason for the failure and then
replace them with others of the same type
There is a serious malfunction; try switching off the control unit
for a few seconds; if the condition persists, it means there is a
malfunction and the circuit board has to be replaced
Control unit works correctly
This is normal when there is a change in one of the inputs: OPEN,
STOP, triggering of photocells or the radio transmitter is used
All OK; when a command arrives, the control unit will reset to
normal functioning (after a brief delay)
An overload has been detected therefore the power to the ECS-
Bus switched off. Check, if necessary, by disconnecting each
device one at a time.
To switch on power to the ECSBus you simply need to give the
command, for example, using the radio transmitter
Action
Check the devices connected to the STOP input
STOP input active
Action
OPEN input not active
Normal only if the device connected to the OPEN input is active
B
A
C
N
M
N
M
FLASH
ECS bus
STOP
OPEN
D
E
F
English – 19

Publicité

Table des Matières
loading

Table des Matières