4.6 COMPLIANCE WITH INTERNATIONAL STANDARDS

Application of safety rules as per standard
EN-61010-1-Ed.2:2001.
Electrical safety :
1000V CAT-III - 600V CAT-IV.
Pollution level 2. Double insulation.
Compliant with standard EN-
Electromagnetic
61326/A2:2001
compatibility :
Residential environment
Mechanical
Free fall: 1m (in accordance with standard
strength :
IEC-68-2-32)
Degree of protection
IP54 as per EN 60529
of the housing :

4.7 VARIATIONS IN DOMAIN OF USE

Quantity of influence Range of influence
Temperature
0° C to + 50°C
Frequency
1 to 3 kHz

5. MAINTENANCE

For maintenance, use only the replacement parts specified.

5.1 CLEANING

Disconnect all leads from the instrument and set the switch
to
.
Use a soft cloth, dampened with soapy water. Rinse with a
damp cloth.
Dry thoroughly with a dry cloth or forced air before using
again.
English
Influence
V AC : 0.5%L/10° C
mV DC : 0.5%L/10° C
(R>20 M ) :
0.5%L/10° C
µF (C>50 µF) :
5%L/10° C