Mounting
Installation of 3A approved and EHEDG certified products:
1) Use only a 3A approved counter part
2) The inspection hole should be visible and drained
3) Mount the adapter in a self drained position/angle
4) Level the inner surface of the pipe with the counter part.
After installation
Check the leak tightness of the sleeve.
Check the tightness of glands or M12 plugs.
Check the tightness of the cover
The hole through the sensor must be in the flow direction
There must be some bi-directional flow through the hole to ensure a
homogeneous media - i.e. correct measurement. There must be no
air-bubbles in the hole.
www.baumer.com
Example with
PM023 or PM053
Example
with PM060-1
Leakage indication hole
must be places downwards
The sensor must be fully submersed in the media.
The marks on the
connection must
point in flow direction
To avoid turbulence problems the sensor is recommended to mount in
a distance of >1 meter from a bend.
CombiLyz, AFI4
Operators Instruction
5) The 3A mark or the arrow shall be placed upwards
6) Welding should be grinded to Ra= 0.8
7) Tighten the swirling nipple on the AFIx 20...25 Nm
8) Installed with correct torque the electrical connection
will point downwards
Operators Instruction: 11126179 00
WARNING
Cannot be
drained
Fitted
incorrect
Example
with PM060-2
Page 5 / 36