Set-up the AFI4
Conductivity measurement is very temperature dependent.
Typical temperature dependence
Media
Acids
Bases
Salts
Neutral water
The factory set-up is 2,0%/°C.
This means that the AFIx will automatically compensate with this fac-
tor. If the temperature dependence is known for the media this factor
must be set. However if the factor is unknown, it is recommended to
use an average from above table.
It is possible to adjust the temperature compensation factor from 0,0 ...
5,0 %/°C.
Configure the temperature compensation by the touch screen:
Configure the range by the touch screen:
Configuration by the FlexProgram and FlexProgrammer please see
HELP in the FlexProgram.
Concentration output can also be selected by the touch screen or pro-
grammed by the FlexProgram.
Programming of the display colours, warnings and relays can be done
by the touch screen (see page 8 and 9) or by the FlexProgrammer.
Selectable display views
Value w. all values
Value w. TAG
Visual alerts and colours
White background
www.baumer.com
% change per °C
1,0 ... 1,6
1,8 ... 2,2
2,2 ... 3,0
2,0
Conductivity
Media w. values
Bar graph w. values
Media w. TAG
Bar graph incl. temp.
Green background
Red background
CombiLyz, AFI4
Operators Instruction
Conductivity ranges (selectable)
0...500 µS/cm
0...1 mS/cm
0...10 mS/cm
0...2 mS/cm
0...20 mS/cm
0...3 mS/cm
0...30 mS/cm
0...5 mS/cm
0...50 mS/cm
Definition:
1.000 µS/cm
=
1.000 mS/cm =
Conductivity in different media:
Conductivity
Media group
55
nS/cm
Water
1
µS/cm
10
µS/cm
100
µS/cm
Food
AFIx
1
mS/cm
range
10
mS/cm
100
mS/cm
Process
To set the external input for range selection
Range
R1
R2
1
N.C.
N.C.
2
24 VDC
N.C.
Concentration ranges (selectable)
NaOH (caustic soda)
0...15% by weight (0...90°C)
25...50% by weight (0...90°C)
HNO
(nitric acid)
0...25% by weight (0...80°C)
3
36...82% by weight (0...80°C)
1 × customer defined (30 point linearization)
Concentration value in %
same views available as for conductivity
Error message and red background
Operators Instruction: 11126179 00
0...100 mS/cm
0...1 S/cm
0...200 mS/cm
0...300 mS/cm
0...500 mS/cm
1.0 mS/cm
1.0 S/cm
Media
Ultra-pure water
Pure water
Process water
Drinking water
Beer
Milk
Orange juice
Apple juice
Phosphoric acid
Hydrochloric acid
Range
R1
R2
3
N.C.
24 VDC
4
24 VDC
24 VDC
Concentration