Component Names
Flow Volume
Adjustment Ring
Nozzle
Nozzle Tip
6

Operation

• Avoid prolonged continuous use, and allow the patient to rinse at appropriate times.
• Never impact the Powder Case and Powder Case Cap as this may cause the Powder
WARNING
Case or Powder Case Cap to scratch or crack.
• Never take the Perio Mate apart during operation. Never pull the Handpiece Release
Collar or the coupling release during operation. Attempting to do this may cause
components of the Perio Mate to unexpectedly detach and the Perio Mate Powder to
scatter.
• Do not open the Powder Case Cap while using the Perio Mate. Powder may scatter.
• Use this product by setting the pressure to 0.3 - 0.4 MPa (3.0 - 4.0 kgf/cm
product at an excessive pressure increases the risk of causing emphysema.
• Thoroughly dry the Perio Mate and all its components inside and out before use.
Insufficient drying may cause clogging of powder.
• Be very careful to avoid Perio Mate Powder entering patient's mucosal areas (eyes,
nose etc) other than the oral cavity. Also, protect the face with a towel, etc., or protective
CAUTION
glasses, etc., to prevent entry of powder into eyes.
• If chemicals, oils, etc, adhere to the Powder Case Cover, wipe them off immediately as
they could cause damage.
Cover Holder
Powder Case Cover
Rubber Packing
Handpiece Joint
O-Ring
Handpiece Release Collar
Powder Case
Handpiece
5
Powder Case Cap
Hose Connection
Internal Nozzle
2
). Using the