6-1 Air Pressure setting
Measure the supply pressure at the handpiece
point and set the pressure to the value specified
on the specification table. (Fig. 1)
*For Multi Gauge information refer to Option Parts
List.
• Do not exceed the optimum pressure specified on the specification table.
• Excess air pressure could cause the Powder Case and Powder Case Cap to crack or
WARNING
fracture, which could lead to injury.
• Make sure the compressed air supply is dry and clean. Water or oil content mixed in the
air may cause internal solidification of the cleaning powder.
CAUTION
6-2 Connecting the Powder Case to the Handpiece Hose
• Before connecting each part, thoroughly dry the Coupling, Handpiece, Hose Joint section
of the Powder Case, inside of the Powder Case, and Nozzle. Also, perform idle running
CAUTION
with air alone to expel any internal water.
Powder Case
AIR
AIR
Powder Case Cap
AIR
Internal Nozzle
Exhaust Air
Pressure
Handpiece
Coupling
AIR
6
Chip Air Pressure
Water Pressure
Drive Air
Pressure
Fig. 1
AIR
Handpiece connection
AIR