5.4.4
Commissioning with vapor
①
Pressure transmitter
②
Stabilizing valve
③
④
,
Differential pressure valves
⑤
Differential pressure lines
⑥
Shut-off valves
Figure 5-2
Measuring steam
Requirement
All valves are closed.
Procedure
WARNING
Hot vapor
Danger of injury or damage to device.
⑥
If the shut-off valves
the pressure transmitter
Follow the specified procedure for commissioning.
•
WARNING
Hot vapor
Danger of injury.
You can briefly open the drain valves
Only open the drain valves
•
SITRANS P DS III/P410 (7MF4.33.. 7MF4.34.. 7MF4.35..)
A5E03434626-04, 05/2015
and the differential pressure valve
①
can be damaged by the flow of vapor.
⑦
to clean the line. Hot vapor can escape in the process.
⑦
briefly, and close them again before vapor escapes.
⑦
Drain valves
⑧
Condensate pots
⑨
Differential pressure transducer/Orifice plate
⑩
Insulation
③
are both open and the stabilizing valve
②
is then opened,
31