7-
Turn/rotate the arms in vertical downwards position for storage.
8-
Secure the vertical and horizontal lock mechanisms.
9-
Hold the support and release the rail clamp.
Note: If the rail clamp is permanently connected to the bar, steps 10 through 12 can be jumped over.
10- Remove the leg support from the rail clamp.
11- Remove the rail clamp from the table rail.
12- Store the rail clamp in an appropriate place.
13- Store the leg support arm in an appropriate place.
14- Repeat step 1 to 13 for the other leg support arm.
XII.
Instruction for Use for Laser Pointer
Class 2 Laser, Laser Diode
Warning: Please note laser radiation is harmful to the eyes, do not look directly into the laser diode or collimated beam
along its optical axis when the diode is activated; and also do not point the laser beam at anyone.
Replacing the Batteries
Remove the battery cover.
Remove the used batteries and replace with three new LR44 / AG13 coin-cell batteries.
Insert cells with the positive (+) side facing out toward the end cap.
Replace the end cap.
Operating the Laser Pointer
Simply press the push button to activate the laser pointer and release it to de-activate.
Warning: Laser pointers are not for children. If not used properly the can be hazardous for the eyes. Do not manipulate
the output of the laser pointer.
No responsibility will be accepted for any damages caused should this security advice be disregarded.
Disposal of Batteries
Batteries must not belong to domestic waste
Dispose the batteries according the local disposals rules.
XIII.
Cleaning and storage of the device
Note: Handle the device with care. The lock mechanisms include safety mechanisms that can be damaged by unprofessional
handling.
1-
After use, clean the surface of the device, with maximum extended tube, with detergent or an antiseptic solution as
per local OR policy/ procedure.
Warning: Do not submerge the device in the detergent or an antiseptic solution, this could damage the locking
mechanisms and may lead to malfunction of the device resulting in severe injury of the patient.
Warning: Do not autoclave.
Note: Only clean the surface. Avoid spilling of detergent or an antiseptic solution in the joint mechanisms. The
mechanisms are semi-closed and provided with grease. Spilling of soft soap and/or alcohol in the mechanisms may
lead to malfunction of the system.
2-
Store the device and its accessories in a dry place.
XIV.
Replacing the straps
Replace straps showing malfunction or if they are contaminated.
Unlock the Velcro on both sides and remove the strap.
Guide a new strap through a hole in the side of the shoe and loop back to secure one side of the strap.
Only use original approved and supplied straps.
XV.
Maintenance of the leg support
The leg support is a mechanical device that needs maintenance as result of wear.
If locking mechanisms do not lock or need extra force to lock, do not use the device until it has been serviced.
The service interval is every two years. Contact your local distributor for service information.
XVI.
Disclaimer of Warranty and Limitation of Remedy
There is no express or implied warranty, including without limitation any implied warranty of merchantability or fitness
for a particular purpose, on the Opitek International product(s), described in this publication. Under no circumstances
shall Opitek International be
87100104 rev 8.0 IFU
Page 6 of 32