5
A4
Frequency Inverter - 3kW, safety class 3
(Safe standstill part only)
D
5V
Inverter for Disc motor
C
A3
Frequency Inverter - 3kW, safety class 3
(Safe standstill part only)
5V
B
Inverter for Sample mover motor
A
5
4
X3.1
K31
RD1
K32
RD2
SB3-3
X20-2
X20-1
34
RD2
RD1
1
2
RD2
X20-3
33
RD1
X3.1
K31
OG
K32
BN
34
RD
33
BK
1
RD
SB3-2
2
OG
EMERGENCY
STOP
1
BN
SB3-1
2
BK
4
3
CPU - BOARD, A1
X107-1
5V
SAFE STANDSTILL FEEDBACK (A4)
X107-11
GND
X107-2
BU6
24V
24V
INVERTER A4 CONTROL
X107-12
BU5
GND
GND
TO MICROPROCESSOR (P63)
X107-9
5V
SAFE STANDSTILL FEEDBACK (A3)
X107-19
TO MICROPROCESSOR (P74)
GND
X107-10
INVERTER A3 CONTROL
X107-20
X114-6
GND
X114-1
24V
24V
X114-7
INTERLOCK CONTROL
GND
TO MICROPROCESSOR (P02)
X114-2
5V
EMERGENCY STOP
3
2
X127-2
X21-2
BK
PSU 24VDC
X127-1
X21-1
BN
X108-4
12
BK4
GND
X108-1
11
BK1
5V
INTERLOCK FEEDBACK
X108-5
22
BK5
32
X108-2
21
BK2
X108-6
A2
BK6
GND
X108-3
A1
BK3
(shown with door open and unlocked)
LOCK ON
Mounting
GND
holes
REV A: OH (2006/10/08)
REV B: OH (2007/01/09)
- S1 and S2 on T1 changed
- cable for YS1 changed
- see changes on other pages
REV C: OH (2007/05/03)
REV D: OH (2007/07/27)
- see changees on page 3
REV E: OHO (2008/11/20)
- see changees on page 3
REV F: TKO (2009/04/02)
- see changes on page 4
Friday, April 17, 2009
2
1
D
T1
RD4
-- 0Vac
S1
RD3
-- 24Vac, 4A
-- NC
F6
-- 0Vac
S2
RD7
-- 24Vac, 4A
Transformer
C
YS1
DOOR
SWITCH
DOOR
SWITCH
B
DOOR
LOCK
Door Guard Switch
A
STRUERS A/S
Pederstrupvej 84
DK-2750 Ballerup
Denmark
ABRAPOL-20, CIRCUIT DIAGRAM
SAFETY RELATED CIRCUIT
Size
CAGE Code
DWG NO
Rev
A2
15883120
F
Scale
Sheet
OH
5
of
5
1