5
QS1
L1
3
L2
L3
3 x FUSE
D
PE
PE
C
LED LAMP
A10
2
B
SUPPLY IN:
4
U1 (24Vac, 4A)
U2 (24Vac, 4A)
Inputs:
Sample mover top position
Removal measurement unit installed
Absolute pressure buttom
Absolute pressure top
Voltage supply 1 (input)
Pump extension board output
Voltage supply 2 (+15V)
Voltage supply 3 (+24V)
Removal measurement
Dosingunit installed
A
5
4
L10
F1-3
3
3
Mains choke
3
4
TO CPU
A3
BOARD, A1
SAFE STANDSTILL
FREQUENCY
INVERTER 3kW
3
TO CPU
BOARD, A1
CAN BUS
3
M1
Sample mover motor
550W
MACHINE CONTROL - BOARD, A2.
22
16
Outputs:
4 x 24VDC air valve:
Sample mover up
Ait outlet 1
Ait outlet 2
Sample mover down
6 x 24VDC pupm motors
1 x 24VAC Shifting valve recirk. / drain
4
3
3
Z1
RFI FILTER
3
4
TO CPU
A4
BOARD, A1
SAFE STANDSTILL
FREQUENCY
INVERTER 3kW
3
TO CPU
BOARD, A1
CAN BUS
3
M2
Disc motor
2,2kW
SUPPLY IN:
U1 (24Vac, 4A)
RS485
2
3
2
CONNECTION TO SAFETY
TRANSFORMER
2MT72034
T1
F4-5
2
2 x FUSE
IEC742 SAFETY
TRANSFORMER
DOOR
INVERTERS
INVERTERS
QUARD
(SAFE
(CAN BUS)
SWITCH
STANDSTILL)
8
4
6
2
CPU - BOARD, A1
6
8
8
START
Incremental
STOP
Keyboard
dial buttom
EMRG. STOP
REV A: OH (2006/10/23)
REV B: OH (2007/03/29)
- L10 added
REV C: OH (2007/07/27)
- 2nd PE terminal added
REV D: OHO (2008/11/20)
- Z1 added
REV E: TKO (2009-05-11) - A10 added
Monday, May 11, 2009
2
1
2
U1: 24Vac, 4A
F6
2
2
U2: 24Vac, 4A
1 x FUSE
3 x SOLENOID
VALVE
COOLI
tap water
(REC. SYSTEM)
disc cooling
water to OP-S
2
6
Display
13
320 x 240
2
BACKLIGHT
3
SERVICE
INTERFACE
(RS232)
STRUERS A/S
Pederstrupvej 84
DK-2750 Ballerup
Denmark
ABRAPOL-20, BLOCK DIAGRAM
Size
CAGE Code
Rev
DWG NO
A2
15883050
Scale
Sheet
OH
1
of
1
1
D
C
B
A
E