Display/Problem
The pulse frequency is
higher or lower than
the average value.
The device switches
off automatically.
The device pumps up
during the measuring
process.
Note: All measured values saved in the ibp TD 3213 can be transferred to a computer. To do this, you will
require a cable for data transfer and suitable software (neither are included in delivery). The data
transfer cable can be obtained from our customer service department on request. An English-
language software is available on the manufacturer's web site at www.taidoc.com.
general information 14
Meaning
You moved, or you talked.
You carried out the measurement
directly after physical exercise.
System requirement.
This is normal.
If the user blood-pressure to be determined is higher than
the output value of the measuring cuff pressure, the device
will pump automatically.
Pressure buildup takes place in stages of 20 mmHg, until the
pressure has been reached which is usable for the measuring
process.
The measuring cuff has not been cor-
rectly located or is not tight enough.
Solution
Do not move during the measurement,
and do not talk.
Do not carry out measurements
immediately after physical exercise.
Press the
-button again and
measure once more.
Make sure the measuring cuff is
correctly located.
35