Operation (cont.)
12. When the warming cycle time is zero,
the visual indicator on the unit will flash
green and the audible indicator (if not
muted) will chime.
13. The warmer will automatically go into a hold cycle for up
to 30 minutes. During the hold cycle, body temperature air
will continue to circulate in the warmer to ensure that the
milk does not cool prior to feeding. The visual indicator will
continue to flash green. The audible (if not muted) will chime
every 5 minutes until hold cycle time expires. Remove the
milk before the hold cycle expires.
14. If the lid is opened during the hold cycle,
the visual indicator will flash green, and
the unit will go into Pause Mode for 60
seconds. The syringe or container must
be placed back in the unit, the lid must
be closed, and the
The cycle will continue until the session
is complete.
Note: If the 60 seconds is exceeded, the unit will stop and reset.
15. Once the hold cycle has expired, the unit will no longer
maintain temperature. It will beep every 15 seconds (if not
muted) until the lid is opened or unit is stopped.
Note: Check elapsed time and refer to your hospital guidelines for
breastmilk handling.
16. Gently mix milk and press stop upon removal.
T
Tip: The warmer was designed to warm milk from refrigerator
temperatures (approximately 4˚C/39.2˚F). If milk contains any ice at
the beginning of the cycle, the final temperature of the milk at the
end of the warming cycle may fall below the target range of the
device.