XT Prime /English
± 1 kV line(s) to
Surge
line(s)
IEC61000-4-5
± 2 kV line(s) to
earth
Voltage dips,
Voltage Dips:
short
i) 100% reduction for 0.5 period,
interruptions and
ii) 100% reduction for 1 period,
voltage variations
iii) 30% reduction for 25/30 period,
on power supply
Voltage Interruptions:
input lines
100% reduction for 250/300 period
IEC61000-4-11
Power frequency
(50/60Hz)
30 A/m
magnetic field
IEC61000-4-8
3 Vrms
0,15 MHz – 80 MHz
Conducted RF
6 Vrms in ISM bands
IEC 61000-4-6
between 0,15 MHz
and 80 MHz
80 % AM at 1 kHz
3 V/m 80 MHz to
2.7 GHz
80 % AM at 1 kHz
Radiated RF EM
385-6000 MHz,
Fields
9-28V/m, 80%
IEC61000-4-3
AM(1kHz) pulse
mode and other
modulation
± 1 kV line(s) to
± 1 kV line(s) to
line(s)
line(s)
100-240 V
30 A/m
30 A/m
3 Vrms
0,15 MHz – 80
MHz
6 Vrms in ISM and
amateur radio
6Vrms
bands between
0,15 MHz and 80
MHz
80 % AM at 1 kHz
10 V/m 80 MHz to
2,7 GHz
80 % AM at 1 kHz
385-6000 MHz,
10V/m
9-28V/m, 80%
AM(1kHz) pulse
mode and other
modulation
Mains power quality
should be that of atypical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user
of this device requires
continued operation
during power mains
interruptions, it is
recommended that the
device be powered from
an uninterruptible power
supply or a battery.
Power frequency
magnetic fields should be
at levels characteristic of
a typical location in a
typical commercial or
hospital environment.
Portable and mobile RF
communications
equipment should be
used no closer to any part
of this device, including
cables, than there
commended separation
distance calculated from
the equation applicable
to the frequency of the
transmitter.
Recommended separation
distance
d
=
150kHz to 80MHz
d
=0.6
80MHz to
800MHz
d
=1.2
800 MHz to 2.7
GHz
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the
transmitter manufacturer
and d is the
recommended separation
distance in meters (m).
Field strengths from fixed
RF transmitters, as
20