8. Troubleshooting
The table below lists troubleshooting solutions for the problems that may happen. If the problem
persists, contact your equipment provider service agent.
P r o b l e m
No display
Display code incorrect
Illuminant under LCD is not
on
Buttons disable
Air delivered is slow
Data can not be copied to
the SD card
P o s s i b l e C a u s e s
1. The power cord is not
connected to the power
socket.
2. LCD failure or controlled PCB
failure.
LCD failure or controlled PCB failure. Contact your equipment provider for
LED failure
Button failure
1.
During ramp time.
2.
Filter is too dirty.
3.
Flow generator failure.
1. SD card is full.
2. SD card is not inserted
correctly.
3. Data on the SD card is
corrupted.
S o l u t i o n s
1. Ensure the power cord is
connected.
2. Contact your equipment provider
for repair.
repair.
Contact your equipment provider for
repair.
Contact your equipment provider for
repair.
1. Check the ramp time setting
2. Change or clean the filter
regularly.
3. Contact your equipment provider
for repair.
1. Ensure the SD card has enough
capacity.
2. Ensure the SD card is inserted into
the slot.
3. Format the card.