TUMESCENCE SUPPLY SETUP
WARNINGS:
•
This system is intended only for use with flexible fluid containers and
fluid bags. Glass containers or bottles may break, and there is a risk of
implosion.
•
Use of bags or containers not approved for this system, or large and/or
lopsided loads, may cause the device to tip over.
1.
Hang the tumescence solution bag from the left hand arm of the
TRIVEX System irrigation mast.
NOTE: Fluid bags must only be hung on the bag holder hooks. Hang only one
bag per hook. Maximum allowable volume is 3 liters.
2.
Open the Tumescence Pump Door.
3.
Move the Tumescence Pump Handle to the open position to allow
loading of the tubing into the Tumescence Pump.
4.
Insert the tumescence solution tubing into the Tumescence Pump, and
snap the yellow tubing fittings into the corresponding yellow brackets.
Ensure that the tubing rests securely in the tubing forks.
CAUTION: Use only the TRIVEX System Illuminator Inflow Tube Set (REF
7209513).
WARNINGS:
•
The Tumescence Pump should not be running while setting up the tubing. Injuries to the operator's hands can occur.
•
Tube sets are provided STERILE and are for single use only. Do not reuse. Do not resterilize. Prior to use, inspect the product package for signs of damage or
tampering. Discard any opened and unused product.
5.
Close the Tumescence Pump using the Tumescence Pump Handle. Make sure the tubing forks do not pierce the tumescence solution tubing.
6.
Connect the tubing to the Tumescence solution bag.
NOTE: The Tumescence Solution tubing should only be connected to the TRIVEX System Illuminator.
7.
Close the Tumescence Pump Door.
RESECTOR HANDPIECE SALINE SUPPLY SETUP
1.
Hang the Saline solution bag from the right hand arm of the TRIVEX System irrigation mast.
2.
Open the Saline Pump Door.
3.
Move the Saline Pump Handle to the open position to allow loading of the tubing into the Saline Pump.
4.
Insert the Saline solution tubing into the Saline Pump, and snap the green tubing fittings into the corresponding green brackets. Ensure that the tubing
rests securely in the tubing forks.
CAUTION: Use only TRIVEX System Resector Kit tube sets (REF 7209514 or 7209515).
WARNING: The Saline pump should not be running while setting up the tubing. Injuries to the operator's hands can occur.
WARNING: Tube sets are provided STERILE and are for single use only. Do not reuse. Do not resterilize. Prior to use, inspect the product package for signs of
damage or tampering. Discard any opened and unused product.
5.
Close the Saline Pump using the Saline Pump Handle. Make sure the tubing forks do not pierce the saline solution tubing.
6.
Connect the tubing to the Saline solution bag.
NOTE: The Saline Solution tubing should only be connected to the Resector Handpiece.
7.
Close the Saline Pump Door.
CONNECTING THE RESECTOR HANDPIECE AND RESECTOR
CAUTION: Make sure the wheels on the TRIVEX System roller base are locked to prevent the system from rolling during setup and use.
1.
Unscrew the protective cap from the connector end of the resector handpiece cable.
2.
Connect the resector handpiece cable to the connection port on the front panel as follows: Orient the handpiece cable connector by aligning the white
double arrows on the connector with the white dot on the connection port. The handpiece cable is fully engaged when the white arrows and the white dot
are almost touching. Make certain the handpiece cable is fully inserted into the connection port.
3.
Press and hold the Run button on the resector handpiece to confirm operation. The resector handpiece motor action can be observed by looking inside the
distal end of the resector handpiece.
4.
Remove the resector from the sterile package and insert the resector into the resector handpiece per the TRIVEX Resector Instructions for Use (R2590).
5.
Press the Run button and observe the resector action to verify that it is properly installed.
6.
Set the window lock position as described in the Operation section of this manual.
Figure 6: Tubing Connections
9