POWER SWITCH

The green rocker switch on the right side is the power ON/OFF switch for the entire system
The switch is illuminated when the system is ON.
The three-prong electrical connector allows connection of the system to any 100 to 240 volt AC (50/60 Hz) source using the power cord supplied with the system.
The system power supply automatically detects the local power standard and adapts the system to that standard

FUSE PANEL

The control unit is protected by dual, 6.3 amp/250 volt time delay fuses mounted on the rear panel to the right of the green power switch
NOTE: If system is turned off for any reason, wait at least 15 seconds before turning power back on.
PREOPERATIVE SETUP FOR USE IN A SURGICAL PROCEDURE
WARNING: Work exclusively with sterile substances, sterile fluids, and sterile accessories.
There are five main steps for preparing the TRIVEX System for use in a surgical procedure:
1.
Power up - turn the TRIVEX System on and check for faults.
2.
TRIVEX System Illuminator Tumescence supply setup - prepare the Tumescence solution and tubing.
3.
Resector Handpiece saline supply setup - prepare the saline irrigation and tubing.
4.
Resector Handpiece setup - connect the Resector Handpiece and verify operation.
5.
Connect the TRIVEX System Illuminator and Tumescence Pump footswitches and verify operation.

POWER UP

CAUTION: Ensure that the available mains voltage matches the data listed on the label attached to the back of the control unit. Incorrect voltage can cause errors
or malfunction, and may destroy the equipment.
WARNING: To prevent electrical shock, connect the power cord to a properly- wired grounding receptacle only.
1.
Plug the unit power cord into the rear panel connector and a grounded AC power source. The control unit power supply automatically detects the local
power standard and adapts the system to that standard.
2.
Push the power switch on the rear panel to the ON ( I ) position.
3.
The TRIVEX System will start up with these defaults:
Tumescence pump flow set to level four which corresponds to approximately 450 ml/min flow. Tumescence pump is OFF and Pump ON/OFF button is lit
steady orange.
Saline pump flow is set to level one. Saline pump is OFF and pump ON/OFF button is lit steady orange.
Resector Mode is set to Oscillate (<>). The Resector Mode button will be lit steady orange if no Resector Handpiece is connected. The button will be lit
steady green if a Resector Handpiece is connected.
Resector Speed is set to 500 rpm.
Lamp is OFF. Lamp ON/OFF button is lit steady orange.
NOTE: If the Resector Mode Select Button is flashing green or orange see "TROUBLESHOOTING".
NOTE: If the Lamp ON/OFF Button is flashing orange see "TROUBLESHOOTING".
Figure 5: Control Unit Rear Panel
1
.
8
2
1
2
.