Description; Front Panel; Keys - Chauvin Arnoux C.A 6250 Notice De Fonctionnement

Microhmmètre
Masquer les pouces Voir aussi pour C.A 6250:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 1
2.1
FRONT PANEL OF THE C.A 6250
-
4 4mm-dia. safety terminals identified as C1, P1, P2 and C2
-
9-way rotary switch :
Off
: instrument power off / setting for charging
I
2500 Ω
: 2500,0 Ω range – measuring current 1 mA
I
250 Ω
: 250,00 Ω range – measuring current 10 mA
I
25 Ω
: 25,000 Ω range – measuring current 100 mA
I
2500 mΩ
: 2500,0 mΩ range – measuring current 1 A
I
250 mΩ
: 250,00 mΩ range – measuring current 10 A
I
25 mΩ
: 25,000 mΩ range – measuring current 10 A
I
5 mΩ
: 5,0000 mΩ range – measuring current 10 A
I
SET-UP
: instrument configuration
I
-
1 yellow START/STOP key: start/stop measurement
-
8 elastomer keys each having a primary function and a secondary function.
-
1 backlit LCD screen
-
1 receptacle for connection to line power to charge the battery
-
1 jack for connection of a Pt100 temperature probe,
-
1 RS 232 serial INTERFACE plug (9 pin contacts) for connection to a PC or to a printer.
2.2

KEYS

8 keys each having a primary function and a secondary function :
Activate the secondary function written in yellow italics below each key. The
symbol appears on the screen.
nd
2
Primary function: before starting the measurement, select the desired measurement
mode: inductive, non-inductive or non-inductive with automatic triggering.
Secondary function: select the metal for the temperature compensation calculation:
METAL
Cu, Al, or Other metal.
Primary function: activate/deactivate the temperature compensation function:
R(
)
θ θ θ θ
calculation of the resistance at a temperature other than the measurement
temperature.
Secondary function: activate/deactivate alarms. The direction and the triggering
ALARM
value (high or low) are adjusted in the SET-UP menu
MEM
Primary function: store the measurement at an address identified by an object
number (OBJ) and a test number (TEST).
Secondary function: retrieve stored data (this function is independent of the setting
MR
of the switch) except in the OFF and SET-UP settings.
%
Primary function: in SET-UP mode, select a function or increment a flashing
parameter.
Secondary function: in SET-UP mode, select a function or decrement a flashing
&
parameter.
Primary function: select the parameter to be modified (in wraparound mode, from
left to right). In SET-UP mode, access the adjustments of a function.
Secondary function: in SET-UP mode, shift the decimal point and select the unit.
I
PRINT
Primary function: print the measurement directly to a serial printer.
Secondary function: print stored data to a serial printer.
PRINT MEM
6
Primary function: activate/deactivate the backlighting of the display unit.
Secondary function: activate and adjust the sound level/deactivate the audible
signal.

2. DESCRIPTION

27

Publicité

Table des Matières
loading

Table des Matières