Alternative 3: End-point charging (Standard version for article number 50100051 and accessory for article number 50100057)
If the end-point charging option is chosen, the lift unit must be fitted with a charger contact (70200062). The charger that
is included with the lift unit is used.
1) Check to ensure that the charging points are set for the charger station; the charger plate is secured with the spring
pins, see Figure 1.
2) Check to ensure that the charger plate springs in and out when pressure is applied to the charger plate, see Figure 2.
3) Guide the lift unit onto the rail so that the plus on the charging point is on the right side, according to Figure 3.
4) Mount charger contacts according to Figure 4. NOTE! Push the charger contact (A) in about 50 mm if only a stop bolt
is used. When an adjustable end-stop is used, the charger contact is pushed in about 90 mm.
5) Affix the accompanying screws for charger contacts, see Figure 5 (B).
6) Connect the red cable to the charger contact on the right side, the charger plate's plus side (positive), and the black
cable on the left side (negative), according to Figure 6.
7) Install the end-stop on the rail.
Spring pin
1
4
6
2
A
+
-
+
+
-
3
5
+
+ Plus
+
B
9