NOTICE
ate
1. Read all of the information in the operation guide and any other literature in the box before operating the unit.
e rate.
2. Stay still, calm and rest for 5 minutes before blood pressure measurement.
nic
3. The cuff should be placed at the same level as your heart.
0
4. During measurement, neither speak nor move your body and arm.
ce),
5. Measuring on same wrist for each measurement.
6. Please always relax at least 1 or 1.5 minutes between measurements to allow the blood circulation in your
tibility
arm to recover. Prolonged over-inflation (cuff pressure exceed 300 mmHg or maintained above15 mmHg
uipment
for longer than 3 minutes) of the bladder may cause ecchymoma of your wrist.
invasive
7. Consult your physician if you have any doubt about below cases:
art 1:
1) The application of the cuff over a wound or inflammation diseases;
art
2) The application of the cuff on any limb where intravascular access or therapy, or an arterio-venous (A-V)
shunt, is present;
3) The application of the cuff on the arm on the side of a mastectomy;
4) Simultaneously used with other monitoring medical equipments on the same limb;
5) Need to check the blood circulation of the user.
8.
This Electronic Sphygmomanometers is designed for adults and should never be used on infants or
young children. Consult your physician or other health care professionals before use on older children.
9. Do not use this unit in a moving vehicle, This may result in erroneous measurement.
10. Blood pressure measurements determined by this monitor are equivalent to those obtained by a trained
observer using the cuff/stethoscope auscultation method, within the limits prescribed by the American
National Standard Institute, Electronic or automated sphygmomanometers.
11. Information regarding potential electromagnetic or other interference between the blood pressure
monitor and other devices together with advice regarding avoidance of such interference please see part
ELECTROMAGNETIC COMPATIBILITY INFORMATION.
12. If Irregular Heartbeat (IHB) brought by common arrhythmias is detected in the procedure of blood
pressure measurement, a signal of
Sphygmomanometers can keep function, but the results may not be accurate, it's suggested that you
consult with your physician for accurate assessment.
There are 2 conditions under which the signal of IHB will be displayed:
1) The coefficient of variation (CV) of pulse period >25%.
2) The difference of adjacent pulse period ≥0.14s, and the number of such pulse takes more than 53
percentage of the total number of pulse.
13. Please do not use the cuff other than supplied by the manufacturer, otherwise it may bring biocompatible
hazard and might result in measurement error.
ensor
will be displayed. Under this condition, the Electronic
5 5