9.4
Calling up messages on the
touch screen
The Messages view shows a history of all previ-
ous messages. Messages are subdivided into the
following categories:
Fault
Unit will no longer function.
When the error has been
remedied, it may be neces-
sary to acknowledge the
error message.
Notice
After acknowledgment the
unit will continue to work,
but only with limited func-
tions.
Note
Important information for
the operator, e.g. about
the current status of the
device.
The unit continues to oper-
ate.
Information
Information for the opera-
tor.
The unit continues to oper-
ate.
Fault-free operation
Tap on Messages.
❯
This displays the message. If there are several
messages, the most current with the highest
priority is displayed first.
For more information about the message, tap
❯
on Help.
2151100018L40 2012V016
10 Correct use of phosphor
storage plates
WARNING
Risk of cross contamination when not
using the barrier envelope or when
using the barrier envelope more than
once
Do not use an phosphor storage plate
❯
without a barrier envelope.
Do not re-use the barrier envelope
❯
(disposable item).
CAUTION
Image data on the phosphor storage
plate is not permanent
The image data is altered by light, natu-
ral X-ray radiation and scattered X-ray
radiation. This impairs the diagnostic
information and clarity.
Read the image data within 30 minutes
❯
of exposure.
Never handle an exposed phosphor
❯
storage plate without the barrier enve-
lope.
Do not subject an exposed phosphor
❯
storage plate to X-ray radiation before
and during the scanning process. Do
not X-ray during the scanning process
if the unit is in the same room as the X-
ray tube.
Usage
EN-
US
33