User Controls
ON / OFF Button
Press once to turn "ON"; Press and hold for one second to turn "OFF".
Audible Alarm Button
Pressing this button will toggle the Inogen One® G2's breath detection
alert on and off:
1. Default Mode. When the Inogen One® G2 is powered
up, the breath detection alert is disabled. The display's mode indication
area is empty in Default Mode.
2. Breath Detection Alert Mode. The Inogen One® G2 will alert with
audible and visual signals for "no breath detected" when this mode is
enabled and no breath has been detected for 60 seconds. The display's
mode indication area will show a bell icon when the alert is enabled.
Flow Setting Control Buttons
Use the – or + flow setting control buttons to select the setting
as shown on the display. There are five settings, from 1 to 5.
Display Backlight Button
Push on; automatically turns off.
4
User Interfaces
Display
This screen displays information regarding flow setting, power
status, battery life and errors. If you would like to change the
language on the Inogen LCD Screen contact Inogen Client
Services.
Indicator Lights
A red light indicates either a change in operating status or a
condition that may need response (alarm). A flashing light is
higher priority than non-flashing.
Audible Signals
An audible signal (beep) indicates either a change in
operating status or a condition that may need response (alarm).
More frequent beeps indicate higher priority conditions.
Input / Output Connections
Particle Filter
The filter must be in place at the intake end of the concentrator
during operation to keep input air clean.
Cannula Nozzle Fitting
The nasal cannula connects to this nozzle for Inogen One® G2
output of oxygenated air.
DC Power In
Connection for power connection from the Universal Power
Supply.
Description of the Inogen One® G2 Oxygen Concentrator Chapter 2
5