Periodical Checks To Be Performed On The Product; Routine Maintenance - Tecno-gaz ORION 60 Notice D'utilisation Et D'entretien

Masquer les pouces Voir aussi pour ORION 60:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

TECNO-GAZ
No.
Period
1
Before using
2
Before using
3
Every 6 months
4
Every 6 months
5
Once a year
Operation and
maintenance manual
Use a flathead screwdriver to increase the braking force, turning the
screws (1) and (2) of the arm brake clockwise.
To increase the braking force at the head, turn the two brake screws (3
and 4) clockwise using a flathead screwdriver.
7.3 Periodical checks to be performed on the
Product
CAUTION – Product electrical check
At the time of start up and after each maintenance job, perform
electrical tests and jobs indicated in the IEC 62353 standard.

7.4 Routine maintenance

CAUTION – Interrupt the power supply before doing any maintenance
CAUTION – Check Product integrity
Make sure there are no pieces or fragments of paint that could
become detached and fall within the operating field. If there are any,
remove them manually.
Make sure the light source protection screens are not damaged. If
they are, contact the Service Personnel and have them replaced.
Check all the Product joints and make sure there are no noises or
squeaks. If there are, lubricate the clutches involved with suitable
grease for industrial use at a service temperature between -30°C and
+ 120°C, type OKS 470 or similar characteristics.
If the Product fails to maintain a regular position, adjust the clutches
as indicated at points 7.1 and 7.2 (arm and clutch adjustment).
Make sure the bar retention screws are tightened properly. Also check
the bar horizontal arm retention screws and the swinging arm screws.
If these are not properly fastened, adequately tighten.
MU_025
Rev.2
Page 18 of 28
jobs
Job
04/02/16
EN

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Orion 80

Table des Matières