Emc Declaration - Tecno-gaz ORION 60 Notice D'utilisation Et D'entretien

Masquer les pouces Voir aussi pour ORION 60:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

TECNO-GAZ
The Product has been tested according to EN60601-1-2 standard to ensure correct
electromagnetic compatibility.
Portable and mobile communication appliances can affect the Product. Other appliances used in
the proximity of the product must also be in conformity with this standard.
The Product has been designed to be used in the electromagnetic environments described below.
The Responsible Organization or Operator is responsible for making sure the Product is used in a
compatible environment.
Immunity test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Operation and
maintenance manual

10 EMC Declaration

CAUTION – Possibility of interferences with nearby appliances
Conformity
Group 1
Class A
Class A
Conforming
MU_025
Electromagnetic environment - directives
The Product only uses RF energy for internal
operation. Consequently its RF emissions are very low
and should not cause any interference to nearby
electronic appliances.
The Product is suitable for use in all environments
except in domestic environments and those directly
connected to a low-voltage public mains supply which
supplies buildings used for domestic purposes, as
long as the following precaution is followed.
Warning: This Product is intended for use by
professional health personnel only. This Product can
cause radio-interference or disturb the operation of
nearby appliances. Measures may have to be taken to
reduce such disturbance, such as Product re-
positioning or shielding of premises.
04/02/16
Rev.2