Battery Test / Current Measurement

5.5 Battery Test

Position measurement function selection switch to 9 V or 1,5 V Battery
Test measurement range.
• Connect the black test lead to the COM socket and the red test lead to the
VΩmA socket.
• Connect test leads to UUT.
• Read the measurement result displayed on the screen.
The batteries are tested with 30 mA (1,5 V) and 10 mA (9V).

5.6 Current Measurement

Ensure that the measurement circuit is not live when connecting the
measurement instrument.
The instruments may only be used in current circuits protected at 16 A
up to a nominal voltage of 600 V.
The nominal cross section of connecting line has to be respected and a safe
connection has to be ensured.
After instrument fuse tripping eliminate the cause for the tripping prior
to fuse replacement.
Current Measurement 200 mA
• Position measurement function selection switch to A 200m= or A 200m~
measurement range.
• Connect the black test lead to the COM socket and the red test lead to the
VΩmA socket.
• Connect test leads to UUT.
• Read the measurement result displayed on the screen.
32