Safety Standards:
IEC / EN 61010-1
IEC / EN 61010-2-032
CAT. III 600V, Pollution Degree 2
Electromagnetic Compatibility Standards
(EMC):
EN 61326-1
Application Standards:
IEC 61557-13: Class 2, ≤ 30A/m
@ In: 3.5 to 600 mA, Fn: 40 to 1k Hz
Valid at best resolution in 6 mA, 60 mA and 600
mA range.
When measuring under the following condition
specified in IEC 61557-13, add additional
specification to accuracy specification.
Intrinsic uncertainty
Designa-
or influence quantity
tion code
Position
E1
Supply voltage
E2
Temperature
E3
Distorted waveform
E9
External low
frequency magnetic
field 15 Hz to 400
E11
Hz according to
IEC 61000-4-8
Load current
E12
Touch current
caused by common
E13
mode voltage
Frequency
E14
Repeatability
E15
12
Additional Spec.
±1% of reading
No additional error
0.1 x (spec. accuracy)
/ °C (<18 °C or >28 °C)
No additional error
Add ±10µA error
per 1uT (magnetic
field)
Add ±6µA error
per 1A (load current)
No additional error
No additional error
No additional error