Eliminating Malfunctions; Maintenance And Cleaning - IKA RV 05 basic Mode D'emploi

Table des Matières

Publicité

Note:
- Please note that the heating process takes some time.
- Only fine tune the pressure/temperature parameters once the
bath temperature has been reached.
- Please bear in mind that the cooler load must not exceed
approx. 60%. This corresponds to condensation on approx. 2/3
of the cooling coil.
- Adjust the parameters gradually when using unknown solvents
or mixtures.
- When using mixtures of substances, first be aware of the low-
boiling solvents and work at a low bath temperature initially.

Eliminating malfunctions

The drive is equipped with an overload protection device that will
turn the device off in the event of a long-term overload (motor
overheating). If the device ie turned off during operation and the
orange glow discharge light on the front of the device lights up,
the drive is overheating.
If the evaporator flask is immersed in the bath, it should be remo-
ved from the heat bath immediately by raising the stand. The drive
should be turned off with the On/Off switch and must be left to
cool off for some time.
Important: The device must be turned off to cool down. It will not
turn back on automatically when the motor has cooled of. The
device will not start to run again until it has been turned on again.
If the cooling off phase was too short, the orange glow discharge
light coming on when the device is turned on again will indicate
that the motore is still too hot.
RV05b 012005

Maintenance and cleaning

The IKA RV 05 basic is maintenance-free. It is subject only to the
natural wear and tear of components and their statistical failure
rate.
When ordering spare parts, please give the manufacturing
number shown on the type plate, the machine type and the
name of the spare part.
Please send in equipment for repair only after it has been clea-
ned and is free from any materials which may constitute a health
hazard. Please tell us if you have used any materials which are
dangerous or which may constitute a health hazard.
Use only cleansing agents which have been approved by IKA to
clean IKA devices.
To remove use:
Dyes
isopropyl alcohol
Construction materials water containing tenside / isopropyl alcohol
Cosmetics
water containing tenside / isopropyl alcohol
Foodstuffs
water containing tenside
Fuels
water containing tenside
For materials which are not listed, please request information
from IKA.
Wear the proper protective gloves during cleaning of the devices.
Electrical devices may not be placed in the cleansing agent for
the purpose of cleaning.
Before using another than the recommended method for clea-
ning or decontamination, the user must ascertain with the manu-
facturer that this method does not destroy the instrument.
13

Publicité

Table des Matières
loading

Table des Matières