Guidance and manufacturer's declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS
Guidance and manufacturer's declaration – electromagnetic immunity
The Spirometer is intended for use in the electromagnetic environment specified below. The customer or the user
of Spirometer should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge
(ESD) IEC 61000-4-2
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
ENGLISH
IEC 60601
Compliance
test level
level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
3A/m
3A/m
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If
floor are covered with synthetic material, the relati-
ve humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
32