6-2 IMPLANT SYSTEM

Surgic Pro Series can memorize 8 different IMPLANT SYSTEMs. Also, you can program 8 steps in one SYSTEM

6-3 Calibration Function

attached handpiece and motor.
NOTICE
Accessories".
6-3-1 Calibration preparation
Attach the Calibration Bur to the handpiece.
6-3-2 Conducting Calibration
CAUTION
Hold the motor in your hand and make sure that "L" is displayed
there is a beep and "donE" appears on the display.
Load calibration ... torque display "H"
Make sure "H" is displayed on the torque display, plug the
calibration bur into the Calibration Load.
Speed calibration
automatically.
... torque display "L"
Calibration Load, so securely
Calibration Load
Calibration Bur
13
Fig.14