3.6.2 The "STOP" key

By the "Stop" key (10) (See figure 27) you can interrupt the run at any time. After pressing the key the centrifuge
decelerates with the respective pre-selected intensity down to stand still.

3.7 Imbalance detection

In case of the rotor not being equally loaded, the drive will turn off during acceleration. The rotor decelerates to
stand still.
When in the display "Time/Prog" (A-3) the word "error" (M11) together with the number "01" appear, the weight
difference of the samples is too large. Distribute the weight evenly.
Load the rotor as described in chapter 3.1.2 and 3.1.3.
When inside the display "Time/Prog" (A-3) the word "error" together with the number "02" (See figure 28) appear,
it could be due to the following reason: The imbalance switch is defective.