6 Maintenance and repair
Care with dental care spray
The medical product must be maintained after each cleaning and before
each sterilization.
1. Remove milling cutter or grinder.
2. Cover medical product with maintenance bag.
3. Place medical product on the cannula and activate spray button for one second.
Collet care
Perform collet care once a week.
Remove milling cutter or grinder and spray with the tip
of the spray nozzle into the opening.
Page 31