5 Operation
Attaching the turbine
NOTE! An incorrectly screwed on turbine can loosen from the supply hose and fall down!
Damage to the turbine.
Check before each treatment, if the turbine is securely attached to the supply hose by pulling carefully.
NOTE! Activation of the foot switch while attaching the turbine!
Damage to the turbine and supply hose.
Do not attach turbine with activated foot switch.
NOTE! Turbine not exactly attached on the coupling!
Inexact coupling can destroy the high-pressure lamp of the coupling device or reduce its life span especially when switch off is delayed.
Check for secure placement in the coupling by pulling on the turbine.
1. Attach the turbine straight on the coupling in axial direction (e.g. CONNEX)
and push until it audibly locks into place.
2. Check for secure placement in the coupling by pulling on the turbine.
The turbine is also compatible with: KaVo MULTIflex® couplings.