5 Operation
Pulling off the turbine
NOTE! Risk by activation of the foot switch while pulling off the turbine!
Damage to the turbine and supply hose.
Do not pull off the turbine with the foot switch activated.
Hold the coupling and pull off the turbine with a slight turn in axial direction.
Attaching the drill
WARNING! Injury by sharp edges!
Infections or cuts.
When testing, attaching and removing wear gloves or other types of finger protection.
1. Firmly press the push-button down with the thumb and at the same time
attach the drill as far as possible into the clamping mechanism.
2. Check for firm position of the drill by pulling.
3. Check the lid for firm position with the finger.