4 Installation and commissioning
To operate the turbine an operating pressure of at least 2.1 bar (30 psi) is necessary!
 Check air pressure.
Pressure values for turbine
Driving air
Recommendation
Return air
Water
Spray air
The air consumption is 39 - 55 Nl/min.

Supply of coolant

CAUTION! Heating of the tooth by insufficient amount of water!
Thermal damage of the dental pulp.
 Set amount of water for spray cooling to at least 50 ml/min.
Page 23
2.1 - 4.2 bar (30 - 60 psi)
2.8 bar (40 psi)
< 0.5 bar (< 7 psi)
0.8 - 2.5 bar (11 - 36 psi)
1.0 - 4.0 bar (14 - 57 psi)