4 Installation and commissioning

4Installation and commissioning
WARNING! Danger from non-sterile products!
Risk of infections for treating personnel and patients.
 Before first commissioning and after each use the medical product must be processed appropriately and if necessary sterilized.
NOTE! Danger during commissioning of the turbine after storage in very cold place!
Possible malfunction of the turbine.
 Bring massively cooled products to a temperature between 20-25°C (68-77°F) before commissioning.
NOTE! Damages by polluted and moist cooling air!
Polluted and moist cooling air can lead to malfunction and premature wear of the bearings.
 Provide dry, clean and uncontaminated cooling air pursuant to ISO 7494-2.
NOTE! Missing or damaged gaskets!
Missing or damaged gaskets lead to premature wear or malfunction. All 5 gaskets must be applied and must not be damaged.
 Check turbine coupling.
Page 22