Changing the electrolyte, the membrane body
or the interior body
METTLER TOLEDO DO sensors are supplied with fitted
membrane body and have been checked for proper
function.
However, if a sensor is to be stored for several months,
the electrolyte should be replaced before use.
If the membrane exhibits signs of failure (long
response time, increased current in an oxygen-free
medium, mechanical damage, etc.) the membrane
body has to be replaced.
Warning! The O
of 13. Contact of electrolyte with mucous membrane
or eyes is to be avoided. Therefore protective gloves
and safety glasses have to be worn for the following
dismantling works.
If such contact occurs, the affected area should be well
rinsed with water. In the case of accident, or should
ever any adverse signs appear, get immediate medical
attention.
When changing the electrolyte, the membrane body or
the interior body, please observe the following instruc-
tions (see also the following illustration):
Attention! Make sure that this maintenance step is
carried out in clean place.
1. Unscrew the cap sleeve from the sensor shaft and
carefully pull it off the sensor.
2. Pull off the membrane body from the interior body.
If it is tight-fitted, eject by pushing it with the flat
finger tip. Before electrolyte is refilled, the mem-
brane body must be removed from the cap sleeve.
3. Rinse the interior body with demineralized water
and carefully dab it dry with a paper tissue.
Note: steps 4 to 7 may only be carried out when
changing the interior body.
4. Unscrew the retainer nut of the interior body with
an adjustable wrench or with a 3/8" wrench.
5. Remove the interior body by pulling it out of the
sensor shaft. If necessary use a plier.
Warning! Do not twist the interior body. Other-
wise the connection pins can be damaged.
InPro 6800
52 200 953 B
InPro 6800 Series O
electrolyte has an alkaline pH value
2
© 12 / 2018 Mettler-Toledo GmbH
Sensor 12/25 mm