When using high intensity illuminators at full output, the recommended
Caution
distance from the headlight module to the patient is not less than 12 inches (30.5 cm). If using
less than 12 inches from the patient, the light intensity must be reduced.
Caution
User is responsible for determining if interruption of light output will create
an unacceptable risk.
illuminator) should be made to reduce the risk.
Caution
Use of this illuminator with accessories or attachments requires that the end-
user follow all accessory or attachment instructions which could affect illuminator setup, usage
and/or settings.
3. SPECIFICATIONS
Item
Lamp type
Power
Color temperature
Lamp life
Lamp replacement
Light guide adapter
Brightness control
Input voltage
Power consumption
Regulatory Approvals
Equipment Class
Mode of Operation
Water Resistant
Operating Environment
Temperature
Relative Humidity
Air Pressure
Storage Environment
Temperature
Relative Humidity
Air Pressure
Dimensions, mm (inch)
Weight, kg (lb)
LIT-219 SUNOPTIC SURGICAL
(English)
If this determination is made arrangements (such as a standby
Specification
Ceramic type 400 watt Xenon
400 Watt
5600º K
1000 hours (typical)
Cartridge replacement from the side of the unit
Turret type to accommodate standard Storz, ACMI, Wolf,
Olympus cable endtips
Mechanical progressive shutter control from the front panel
100-120 V AC, 50/60 Hz
220-240 V AC, 50 Hz
750 watt max
UL60601, EN 60601-1, EN 60601-1-2, CAN/CSA, C22.2,
No 601.1-M90
BF-type
Continuous operation
Not Protected Equipment, IPX0
+20° to +40° C (68° to 104° F)
30 to 85%
700 to 1060 hPa
-20° to +60° C (-4° to 140° F)
0 to 95%
700 to 1060 hPa
338 (13.3)Width x 155 (6.1)Height x 457 (18)Depth front-to-back
10kg (22lb)
®
Rev. -
Date of Revision: 06/21/16
Page 5 of 104