Sunoptic Surgical Titan x450 Mode D'emploi page 10

Masquer les pouces Voir aussi pour Titan x450:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

6.2.1.1
(Refer to 8.1 LAMP REPLACEMENT) Disconnect power. Open the door. The lamp
positioning lever should be completely down(horizontal) to properly engage the lamp
and completely up (vertical) to release it. Raise the lever completely if not already
raised and firmly push the lamp to assure full engagement to the power connectors.
Lower the lamp engagement lever counter-clockwise, from vertical to horizontal, and
observe that the lamp rises slightly as it is engaged into proper operating position.
6.2.1.2
6.2.1.3
The Tx450D digital control system communicates with Sunoptic Surgical lamps designed
for it. Only Sunoptic Surgical lamps can be engaged by the positioning mechanism.
LIT-219 SUNOPTIC SURGICAL
(English)
ENGAGE LAMP
CLOSE DOOR
Close the door completely. (Fan will continue to run with door open.)
INCOMPATIBLE LAMP MODULE INSTALLED
Install correct SSX0450 lamp.
®
Rev. -
Date of Revision: 06/21/16
Page 10 of 104

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Tx450d

Table des Matières