The operator shall not touch the ther-
mometer and the patient simultaneously.
Keep the machine out of children's reach.
* For accidental swallow of bat-
tery or protective film, please con-
sult the doctor at once.
For temperature difference between stor-
age area and measurement site, condi-
tion the thermometer for about 30 minutes
in room temperature (measurement site).
* Failing to do so may cause inaccuracy.
Ensure that children do not use
the instrument unsupervised.
* Some parts are small enough
to be swallowed.
ERROR MESSAGE
ERROR
PROBLEM
MESSAGE
Tempera-
ture taken
is higher
than 42.2°C
Hi
(108°F).
Tempera-
ture taken
is lower
than 32°C
Lo
(89.6°F).
Sensor not
Err
welded
EEPROM is
ErE
abnormal
Low power
Temperature
of the
ErH
measuring
environment
is too high.
Temperature
of the
ErL
measuring
environment
is too low.
MAINTENANCE, STORAGE
AND CALIBRATION
Maintenance
Remove any stains on the body
1
with a soft and dry cloth.
Gently wipe the dirt on the body with a soft cloth,
2
use a cotton swab or soft cloth stained with alcohol,
and gently wipe the lens of the temperature mea-
suring head.After using the product, the product
can be sterilized with 75% medical alcohol on the
WARNING
SOLUTION
Operate the ther-
mometer only between
the specified temper-
ature ranges. In the
event of a repeated
error message, con-
tact your retailer
or Customer Services
Operate the ther-
mometer only between
the specified temper-
ature ranges. In the
event of a repeated
error message, con-
tact your retailer
or Customer Services
Contact your
retailer or Cus-
tomer Services
Contact your
retailer or Cus-
tomer Services
Replace the
batteries
Decrease the
temperature of the
environment, keep
it within 16°C-35°C.
Increase the
temperature of the
environment, keep
it within 16°C-35°C.
soft towel or cotton sliver.
If the probe tip is damaged, please
contact the maintenance center.
Don't wash with water or detergent
3
containing abrasive or benzene.
Do not immerse in liquids.
4
Storage
The device must not be stored or used at an exces-
1
sively high or low temperature or humidity, in sun-
light, in association with an electrical current or in
dusty locations. Otherwise may occur inaccuracies.
Remove the battery if non-use for long time.
2
Calibration
This thermometer is calibrated at the time of manufactured.
If the thermometer is used according to the instruction,
periodic recalibration is not required. If at any time you
question the accuracy of the measurement, please contact
the retailer immediately. Don't attempt to modify or
reassemble the thermometer.
PARAMETER
SPECIFICATION
Measure-
Forehead mode
ment Mode
Measur-
Forehead
ing site
Measure-
Celsius (°C) and Fahr-
ment Units
enheit (°F)
Resolution
0.1 (°C/°F)
Operating
16°C~35°C (60.8°F~95°F) with a
Conditions
relative humidity of 15%~80%
Storage Con-
-20°C~+55°C (-4°F~131°F) with a
ditions
relative humidity of 15%~93%
Measurement
15mm to 50mm from the
distance
measuring point
Forehead
Measurement range:
temperature
32.0°C~42.2°C (89.6°F~108°F)
measurement
Measurement accuracy (Labra-
range and
tory) ±0.3°C (32.0~-42.2°C)
accuracy
/ ±0.5°F (89.6°F~108°F)
Memories
32 measurements recall
recall
Dimensions
138mm × 37mm × 37mm
Weight
47g without batteries
Power
2 × AAA batteries (DC 3V)
Automatically switches
Switch off
off after 1 minute
Product Life 5 years
New bat-
1000 times
tery life
STANDARD
The product is made under the ISO80601-2-56.
GUARANTEE
Two-year warranty is available from purchasing date,
excluding user-caused failures listed below: