Before Use; During Use; Safety Features - Metrix MX 23 Notice De Fonctionnement

Masquer les pouces Voir aussi pour MX 23:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 4
Definition of installation categories :
• CAT IV:
Measurement category IV corresponds to measurements taken at the
source of low-voltage installations.
Example: power feeders
• CAT III:
Measurement category III corresponds to measurements on building
installations.
Example: industrial machinery or instrument power supplies
• CAT II:
Measurement category II corresponds to measurements taken on
circuits directly connected to low-voltage installations.
Example: household appliance and portable tool power supplies
• CAT I:
Measurement category I corresponds to measurements taken on
circuits not directly connected to the network.
Example: protected electronic circuits

1.2.1. Before use

• Comply with environmental and storage conditions.
• For your safety, before using the leads, systematically check they are is in
perfect working condition.

1.2.2. During use

As a safety measure, use only suitable leads and accessories supplied
with the instrument or approved by the manufacturer.
If the measurement category of the accessory is different from that of the
apparatus, the lowest category applies to the unit.
The probes used for network voltage measurement must have at least one
measurement category III according to IEC 61010-031 assigned and rated
voltage less than the voltage of the circuit to be measured.
Never exceed the protection limit values indicated in the specifications for
each type of measurement.
Before changing the function, disconnect the measurement leads from the
circuit measured.
Never measure resistances on a live circuit.
When the instrument is connected to the measurement circuits, never touch an
unused terminal.
1.3.

Safety features

It is impossible to access the battery or fuse compartment without first
disconnecting the measurement leads.
When measuring voltages greater than 50 V
flashes on the display.
When measuring voltages above 7 A, the sign blinks
an intermittent audible signal warns the user.
When there is a persistent overload on following ranges, an intermittent buzzer
indicates the risk of electric shock : VLOW Z, V
500 mA
(or 50 mA
DC
(or 50 mA
AC+DC
(except MX 23).
18
or 30 V
DC
AC+DC
: MX 24 and except MX 23), 500 mA
DC
: MX 24 and and except MX 23), 10 A
Chapter II
, the
symbol
AC
on the display and
, V
, ADP
, ADP
,
DC
DC
AC+DC
AC+DC
, 10 A
DC
AC+DC
Portable digital multimeters

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Mx 24Mx 24b

Table des Matières