The _85RF/_85RF bridge
being switched on and all components being con-
nected properly
Disconnect the mains supply unit from the mains out-
let. Wait a minute and then reconnect it. If necessary,
connect a bedside lamp to check that mains voltage
(240 V) is present. If there is no current, check the fuse.
Press the "all motors down" button (7) to reset all ad-
justments. The control of the _85RF/_85RF bridge
contains microprocessors, so irregular power supply
can lead to malfunctions.
The radio remote control does not work (The back-
ground light
does not work)
Insert new batteries. Where applicable, remove any
obstacles between the radio remote control and the
motor control.
After a power failure once power has been restored
to the system
To restart the initial setup of the _85RF/_85RF bridge
press button (30) on the switching power supply (28)
until all of the motors have powered down and this is
confirmed by the motor's emergency light blinking twi-
ce. (Can take 20 seconds).
Problem
Possible cause
Remote control
No mains supply.
or control system
Manual switch or control system
not working.
defective.
Control systems
Thermoswitch on the transformer
suddenly no
may have been triggered or may be
longer travel /
defective.
move.
Thermal cut-out in the transformer
may have been triggered or may be
defective.
Internal fuse may have been triggered
or may be defective.
No mains supply.
Supply line (mains and / or manual
switch / auxiliary control system) inter-
rupted.
The battery-
Block battery / block batteries flat.
operated electric
reset function
Block battery/batteries not connected.
cannot be acti-
vated.
24
Operating instructions
does not work despite
®
®
,
®
Technical data
Supply voltage: 110-240 V
Remote control power supply: 3 x AAA Batteries
Power rating motors:
Intermittent operation FROM 2 min / 18 min
Power rating user socket : 2000 W / 10 A
Thrust motors: 2 x 6000 N
Power supply for emergency lowering:
2 x 9V block battery
Solution
Restore mains connection.
Contact your supplier / stockist.
Leave the control system in the rest position for 20-30
minutes.
Contact your supplier / stockist.
Contact your supplier / stockist.
Restore mains connection.
Check the supply line and restore contact if applicable.
Check and replace the block battery / block batteries if
necessary.
Connect the block battery / block batteries.
EN